Campo DC | Valor | Idioma |
dc.contributor.author | Dias, Fábio de Souza | - |
dc.contributor.author | Bonsucesso, Josemário Santana | - |
dc.contributor.author | Alves, Lucylia Suzart | - |
dc.contributor.author | Silva Filho, Domingos Correia da | - |
dc.contributor.author | Costa, Antonio Celso Spinola | - |
dc.contributor.author | Santos, Walter Nei Lopes dos | - |
dc.creator | Dias, Fábio de Souza | - |
dc.creator | Bonsucesso, Josemário Santana | - |
dc.creator | Alves, Lucylia Suzart | - |
dc.creator | Silva Filho, Domingos Correia da | - |
dc.creator | Costa, Antonio Celso Spinola | - |
dc.creator | Santos, Walter Nei Lopes dos | - |
dc.date.accessioned | 2013-07-10T18:43:10Z | - |
dc.date.available | 2013-07-10T18:43:10Z | - |
dc.date.issued | 2013 | - |
dc.identifier.issn | 0026-265X | - |
dc.identifier.uri | http://www.repositorio.ufba.br/ri/handle/ri/12165 | - |
dc.description | Texto completo. Acesso restrito. p. 363–367 | pt_BR |
dc.description.abstract | In the present work, a minicolumn of sisal fiber loaded with tiazolylazo-Resorsinol (TAR) is proposed a
preconcentration system for cadmium determination in drinking water samples by flame atomic absorption
spectrometry. In the optimizationwas first of all a full factorial design of two levels (24)was used for preliminary
evaluation of four factors, involving the variables: sampling flow rate, elution flow rate, buffer concentration
and pH. This design has showed that, for the studied levels, buffer concentration and pH are significant factors.
Using the established experimental conditions in the optimization step of: pH 7.0, buffer concentration of
0.009 mol L−1 for elution utilized HCl at 2.0 mol L−1, this system allows the determination of cadmium with a
detection limit (LD) (3σ/S) of 0.05 μg L−1 and a quantification limit (LQ) (10σ/S) of 0.17 μg L−1, precision
expressed as a relative standard deviation (R.S.D.) of 2.9 and 3.4% for a cadmium concentration of 10.0 and
20.0 μg L−1, respectively, and a preconcentration factor of 30 for a sample volume of 50.0 mL. The accuracy
was confirmed by cadmium with spike tests with recuperations varying from 92% to 103%, procedure was
applied for cadmium determination in drinking water samples collected from Salvador City, Bahia, Brazil.
From the five samples analyzed, the achieved concentrations varied from 0.10 to 0.54 μg L−1. | pt_BR |
dc.language.iso | en | pt_BR |
dc.publisher | Microchemical Journal | pt_BR |
dc.source | http://dx.doi.org/10.1016/j.microc.2012.01.018 | pt_BR |
dc.subject | Sisal fiber | pt_BR |
dc.subject | Preconcentration | pt_BR |
dc.subject | Cadmium | pt_BR |
dc.subject | TAR | pt_BR |
dc.title | Development and optimization of analytical method for the determination of cadmium from mineral water samples by off-line solid phase extraction system using sisal fiber loaded TAR by FAAS | pt_BR |
dc.title.alternative | Microchemical Journal | pt_BR |
dc.type | Artigo de Periódico | pt_BR |
dc.description.localpub | Salvador | pt_BR |
dc.identifier.number | v. 106 | pt_BR |
Aparece nas coleções: | Artigo Publicado em Periódico (Química)
|