Título: Unusual calibration curves observed for iron using high-resolution continuum source graphite furnace atomic absorption spectrometry
Título(s) alternativo(s): Spectrochimica Acta Part B: Atomic Spectroscopy
Autor(es): Welz, Bernhard
Santos, Lisia Maria Gobbo dos
Araujo, Rennan Geovanny Oliveira
Jacob, Silvana do C.
Vale, Maria Goreti R.
Okruss, Michael
Ross, Helmut Becker
Autor(es): Welz, Bernhard
Santos, Lisia Maria Gobbo dos
Araujo, Rennan Geovanny Oliveira
Jacob, Silvana do C.
Vale, Maria Goreti R.
Okruss, Michael
Ross, Helmut Becker
Abstract: The simultaneous determination of cadmium and iron in plant and soil samples has been investigated using high-resolution continuum source graphite furnace atomic absorption spectrometry. The primary cadmium resonance line at 228.802 nm and an adjacent secondary iron line at 228.726 nm, which is within the spectral interval covered by the charge-coupled device (CCD) array detector, have been used for the investigations. Due to the very high iron content in most of the soil samples the possibility has been investigated to reduce the sensitivity and extend the working range by using side pixels for measurement at the line wings instead of the line core. It has been found that the calibration curves measured at all the analytically useful pixels of this line consisted of two linear parts with distinctly different slopes. This effect has been independent of the positioning of the wavelength, i.e., if the Cd line or the Fe line was in the center of the CCD array. The most likely explanation for this unusual behavior is a significant difference between the instrument width ΔλInstr and the absorption line width ΔλAbs, which is quite pronounced in the case of Fe. Using both parts of the calibration curves and simultaneous measurement at the line center and at the wings made it possible to extend the working range for the iron determination to more than three orders of magnitude.
Palavras-chave: High-resolution continuum source AAS
Electrothermal atomization
Iron determination
Calibration graphs
Line wings
Editora / Evento / Instituição: Elsevier
URI: http://www.repositorio.ufba.br/ri/handle/ri/5980
Data do documento: 2010
Aparece nas coleções:Artigo Publicado em Periódico (Química)

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
(36)1-s2.0-S058485471000008X-main.pdf
  Restricted Access
250,68 kBAdobe PDFVisualizar/Abrir Solicitar uma cópia
Mostrar registro completo do item Visualizar estatísticas


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.